Q315r manual
last modified
13-09-2006 11:24
System Startup and Shutdown Summary
Startup Summary:
a. Vacuum under 10 microns, both meters.
b. Coolant flowing and at 12 degrees C.
c. Frame grabber computers on
d. Bottom detector power supply on, wait 5 seconds, middle detector power supply on, wait 5 seconds, top detector power supply on
e. Bottom TE cooling power switch on, middle TE on, top TE on.
Cooling Summary:
a. Double click the "Operate Detector" on each frame grabber.
b. Double click the "Quantum Console" icon on one frame grabber.
c. Click the "Connect to Detectors" button.
d. Click the "Enable Temperature Control".
e. Using the Temperature set box, put in "0" (do not hit enter or return). Push the "Set Temperature" button. This will lower the temperature to zero degrees C. o When all module status boxes read "Ready at Final Temperature", you are ready to continue.
f. Push the "Ramp to Cold Operating Temperature" button. o When all module status boxes read "Ready at Final Temperature", you are ready to operate the detector.
g Push the "Exit to Operate Detector" button and wait for the Quantum Console to exit.
h. If you use Remote Detector OP, exit Operate Detector and start Remote Detector OP on each computer.
Shutdown Summary:
a. Warm detector to +10 degrees C. (a-d above, push Warm up to +10 C button, then steps g, & h when all modules are at +10 C)
b. Exit all "Operate Detector" windows you may have open.
c. Detector cooling power may be turned off.
d. Detector power may be turned off.
e. Coolant flow may be turned off if desired.
f. Vacuum pump may be turned off IF and ONLY IF the detector is VALVED OFF.