ID29 Device list - EH

last modified 25-02-2008 10:19


The following pages will give you all informations useful to know electronics parts of Experimental Hutch EH.
There are all Experimental elements, items and their electronic control.

VME CRATE ID292

Control Resources    
Name Comments
Number
Type Conventions:
STM = Stepper Motor   21 VPAP/DPAP1 or DPAP2 X = beam Axis; X+ = direction of photons

STM Total
  21   Z = Vertical Axis; Z+ = Up
 
DCME = DC Motor with encoder   1 Galil DC motion controler  
DCM Total        
 
LE = Linear encoder CC133 counter VME card 1 Heidenhain - Detector translation  
RE = Rotary encoder CC133 counter VME card 2 Heidenhain - Detector Tilt  
Encoders Total   3    
 
DIN = Digital Input undefined Opto-isolated Input relay 14 STB116/IDC modules & ICV196 VME card  
DIN Total   14    
         
DOUT = Digital Output undefined Opto-isolated Output relay 14 STB116/ODC modules & ICV196 VME card  
DOUT Total   14    
 
Analog Input ADC - Current loop mode 1 ICV150-BCU VME card & STB50 interface  
AIN Total   1    
         
Analog Output  DAC- Current loop mode 0 ICV712-BCU VME card & STB50 interface  
AOUT Total   0    
 
Frequency counter  Counter & Synchronisation 2 VCT6 VME card  
CNT= Frequency counter   2    
Photo of Exp VME Crate
crate
Photo of Rack Motor Driver
exp rack

 


 Logical Name  
SPEC
Name
 Component  
Control
Resource
 Cables
Electronic Control
VPAP RPAP8 DPAP
Chan/Numb
DPAP
Current setting
1 - Vacuum pipe.  
VacumIonPumpController   Ion pump control   Ion Pump Cable: C/0018 Varian controler    
VacumIonPumpTSX17       TSX17 + VME in CC0 (To Front End TSX17 control PLC)      
Mono VacuumGauge   Gauge controler   Cable Penning: C/0023 ; Cable Pirani: C/0028 Vaccum Balzers gauge controler    

 
Photo ExpVaccum Equipment
2 - Diffractometer Table.  
FrontLeg-Z-Translation tvf Stepper motor STM Motor Cable: C/0112 VPAP RPAP8#1 Current Nor/Boost
FrontLeg-Z-Translation   Limit   Limit Cable: C/0117 #M1/0 DPAP2#1 I= 2.3 A switch:E
RightLeg-Z-Translation tvin Stepper motor STM Motor Cable: C/0113 VPAP RPAP8#1 Current Nor/Boost
RightLeg-Z-Translation   Limit   Limit Cable: C/0118 #M2/0 DPAP2#2 I= 2.3 A switch:E
LeftLeg-Z-Translation tvout Stepper motor STM Motor Cable: C/0114 VPAP RPAP8#1 Current Nor/Boost
LeftLeg-Z-Translation   Limit   Limit Cable: C/0119 #M3/0 DPAP2#3 I= 2.3 A switch:E
LeftLeg-Y-Translation trot Stepper motor STM Motor Cable: C/0115 VPAP RPAP8#1 Current Nor/Boost
LeftLeg-Y-Translation   Limit   Limit Cable: C/0120 #M4/0 DPAP2#4 I= 2 A switch:E
FrontLeg-Y-Translation ttrans Stepper motor STM Motor Cable: C/0116 VPAP RPAP8#1 Current Nor/Boost
FrontLeg-Y-Translation   Limit   Limit Cable: C/0121 #M5/0 DPAP2#5 I= 2 A switch:E

 Logical Name  
SPEC
Name
 Component  
Control
Resource
 Cables
Electronic Control
VPAP RPAP8 DPAP
Chan/Numb
DPAP
Current setting
4 - Quadrant photodiode  
InputBeamIntensity + Position ct OT302 controler Serial Line SL Cable: Patch panel S/0111  IBAM/RSL18 ch: I1    
 
5 - Slits Set  
Vertical Gap Slit 1 s1v stepper motor STM Motor Cable: C/0122 VPAP RPAP8#2 Current Nor/Boost
Vertical Gap Slit 1   Limit   Limit Cable: C/0126 #M1/1 DPAP1#1 I= 0.3 A; Switch: B
Vertical Translation Slit 1 t1v stepper motor STM Motor Cable: C/0123 VPAP RPAP8#2 Current Nor/Boost
Vertical Translation Slit 1   limit   Limit Cable: C/0127 #M2/1 DPAP1#2 I= 0.3 A; Switch: B
Horizontal Gap Slit 1 s1h stepper motor STM Motor Cable: C/0124 VPAP RPAP8#2 Current Nor/Boost
Horizontal Gap Slit 1   Limit   Limit Cable: C/0128 #M3/1 DPAP1#3 I= 0.3 A; Switch: B
Horizontal Translation Slit 1 t1h stepper motor STM Motor Cable: C/0125 VPAP RPAP8#2 Current Nor/Boost
Horizontal Translation Slit 1   limit   Limit Cable: C/0129 #M4/1 DPAP1#4 I= 0.3 A; Switch: B
Vertical Gap Slit 2 s2v stepper motor STM Motor Cable: C/0130 VPAP RPAP8#2 Current Nor/Boost
Vertical Gap Slit 2   Limit   Limit Cable: C/0134 #M5/1 DPAP1#5 I= 0.3 A; Switch: B
Vertical Translation Slit 2 t2v stepper motor STM Motor Cable: C/0131 VPAP RPAP8#2 Current Nor/Boost
Vertical Translation Slit 2   limit   Limit Cable: C/0135 #M6/1 DPAP1#6 I= 0.3 A; Switch: B
Horizontal Gap Slit 2 s2h stepper motor STM Motor Cable: C/0132 VPAP RPAP8#2 Current Nor/Boost
Horizontal Gap Slit 2   Limit   Limit Cable: C/0136 #M7/1 DPAP1#7 I= 0.3 A; Switch: B
Horizontal Translation Slit 2 t2h stepper motor STM Motor Cable: C/0133 VPAP RPAP8#2 Current Nor/Boost
Horizontal Translation Slit 2   limit   Limit Cable: C/0137 #M8/1 DPAP1#8 I= 0.3 A; Switch: B
Photo of Slits Set Box
Photo of  Slits Set Box
 Logical Name  
SPEC
Name
 Component  
Control
Resource
 Cables
Electronic Control
VPAP RPAP8 DPAP
Chan/Numb
DPAP
Current setting
6 - Photodiode Io_1, Io_2  
Beam intensity Io_1 I0_1 Vct6 counter CNT Electrometer Novelec I/F ; VCT6 counter VME Ch: 3/1    
        Cable Electrometer/VCT6: S/0      
Beam intensity Io_2 I0_2 Vct6 counter CNT Electrometer Novelec I/F ; VCT6 counter VME Ch: 4/1     
        Cable Electrometer/VCT6: S/0      
Electrometer Gain Io_1 & Io_2 elmeter* Digital Output DOUT ICV196 VME card & STB116 Relay Out Chanel ICV196 Io_1 Chan H:66 M:65 L:64     
High-Middle-Low gain       Cable Electrometer Gain: S/0071 Chanel ICV196 Io_2 Chan H:69 M:68 L:67     
Photo of Slts Box Rear Panel
7 - MS-Shutter  
MS-Shutter local control prodc Digital Output DOUT local control by stb116 & ICV196 Ch:70    
        Cable control : S/0056      
MS-Shutter Remote control prodc Relay Output DOUT Remote control from BAC24 (Spindle control)      
BAC24 Spindle axis Counter   Counter CNT Serial Line Cable control : patch panel S/0056 IBAM/RSL18 ch: i2    
Photo of Spindle Axis Control
Photo of Sample Environment
 Logical Name  
SPEC
Name
 Component  
Control
Resource
 Cables
Electronic Control
VPAP RPAP8 DPAP
Chan/Numb
DPAP
Current setting
8 - PHI Axis.  
PHIRotation phi DC motor + encoder DCME Motion control for DC servomotor DMC-1411 for PC/104 from Gallil    
PHI-Motor encoder   Rotary Encoder RE Aux Encoder on DC motor Incremental Encoder from RV120CC MC    
PHI-MainAxis Encoder   Rotary Encoder RE Main Encoder on rotary Table; Phi Axis Incremental Encoder RON785 Heidenhain    
Z-Translation z stepper motor STM Motor Cable: C/0142 VPAP RPAP8#3 Current Nor/Boost
Z-Translation   Limit   Limit Cable: C/0145 #M3/2 DPAP1#3 I=1.2 A; switch: E
 
9 - Beam stop.  
VerticalPosition  bv Stepper motor STM Motor Cable: C/0146 VPAP RPAP8#3 Current Nor/Boost
VerticalPosition   Limit   Limit Cable: C/0150 #M1/2 DPAP1#7 I= 0.25 A; switch:7
HorizontalPosition  bh Stepper motor STM Motor Cable: C/0147 VPAP RPAP8#3 Current Nor/Boost
HorizontalPosition   Limit   Limit Cable: C/0151 #M2/2 DPAP1#8 I= 0.25 A; switch:7
Pneumatic-translation bsin/out Relay output DOUT ICV196 VME card & STB116 Relay Out Chanel ICV196 Beam-stop inserted :78    
Electrovalve Beamstop        Cable beamstop: Spare cable S/0119 Chanel ICV196 Beam-stop out :77    
Captor of position   Relay input DIN ICV196 VME card & STB116 Relay IN Chanel ICV196 Beam-stop inserted :39    
        Cable beamstop: S/0061 Chanel ICV196 Beam-stop out :40    
Photo of Sample Environment
 Logical Name  
SPEC
Name
 Component  
Control
Resource
 Cables
Electronic Control
VPAP RPAP8 DPAP
Chan/Numb
DPAP
Current setting
10 - Fluorescence Detector  
Motor-Translation Fluo-det Stepper motor STM Motor Cable: C/0152 VPAP RPAP8#2 Current Nor/Boost
Motor-Translation   Limit   Limit Cable: C/0153 #M4/2 DPAP1#4 I= 0.25 A; switch:7
Pneumatic-translation euin/out Relay output DOUT ICV196 VME card & STB116 Relay Out Chanel ICV196 Detector inserted :75    
Electrovalve Detector        Cable Fluo Det: Spare cable S/0119 Chanel ICV196 Detector out :76    
 
11 - Light Source.  
Pneumatic-translation lin/out Relay output DOUT ICV196 VME card & STB116 Relay Out Chanel ICV196 light inserted :74    
Electrovalve Light        Cable Light: Spare cable S/0119 Chanel ICV196 Light out :73    
Photo of Button Box
 Logical Name  
SPEC
Name
 Component  
Control
Resource
 Cables
Electronic Control
VPAP RPAP8 DPAP
Chan/Numb
DPAP
Current setting
12 - Button Box
Button box Enable btnbox Relay output DOUT ICV196 VME card & STB116 Relay Out Chanel ICV196 Btn Box enable :72    
  or prodc     Cable btnbox: Spare cable S/0119      
Phi +90 deg button  Relay input DIN ICV196 VME card & STB116 Relay IN Chanel ICV196 Btn Box +90deg :32    
  (+90deg)     Cable Electrometer Gain: S/0058      
Phi -90 deg button  Relay input DIN ICV196 VME card & STB116 Relay IN Chanel ICV196 Btn Box -90deg :33    
  (-90deg)     Cable Electrometer Gain: S/0058      
Phi 180 deg button  Relay input DIN ICV196 VME card & STB116 Relay IN Chanel ICV196 Btn Box 180deg :34    
  (180deg)     Cable Electrometer Gain: S/0058      
Phi "Home position" button  Relay input DIN ICV196 VME card & STB116 Relay IN Chanel ICV196     
  home     Cable Electrometer Gain: S/0061 Btn box "home position":44    
Det ADSC Backward button  Relay input DIN ICV196 VME card & STB116 Relay IN Chanel ICV196     
  det-B     Cable Electrometer Gain: S/0058 Btn Box Detector Backward :36    
Det ADSC Forward button  Relay input DIN ICV196 VME card & STB116 Relay IN Chanel ICV196     
  det-F     Cable Electrometer Gain: S/0058 Btn Box Detector Forward :35    
Light Source Inserted button  Relay input DIN ICV196 VME card & STB116 Relay IN Chanel ICV196 Light inserted :38    
  light-in     Cable Electrometer Gain: S/0058      
Light Source Out button  Relay input DIN ICV196 VME card & STB116 Relay IN Chanel ICV196 Light out :37     
  light-out     Cable Electrometer Gain: S/0058      
Z-translation "local mode" btnbox Relay output DOUT ICV196 VME card & STB116 Relay Out Chanel ICV196     
  or prodc     Cable Local/remote Z-trans :  Z-local mode DPAP enable :72    
Z-translation Speed potentiometer Rot-pot Potentiometer/DPAP   Cable Z-Speed :  Speed of Z-translation    
Z- Motor Direction of Z trans button  Button/DPAP   Cable Z-Motor Direction :  Direction of Z translation    
  Z+/Z-            
Photo of CCD Detector
Photo of Rack CCD Detector
 Logical Name  
SPEC
Name
 Component  
Control
Resource
 Cables
Electronic Control
VPAP RPAP8 DPAP
Chan/Numb
DPAP
Current setting
13 - CCD Detector  
Detector Xtal-translation det-X stepper motor STM Motor Cable: C/0138 VPAP RPAP8#1 Current Nor/Boost
    Limit   Limit Cable: C/0140 #M6/0 DPAP2#6 I= 1.6 A; switch: A
    Linear encoder LE Encoder Cable: S/ CC133 VME Card   #Gr A / Ch 1
Detector Xtal-Back tilt detB stepper motor STM Motor Cable: C/0138 VPAP RPAP8#1 Current Nor/Boost
    Limit   Limit Cable: C/0140 #M6/0 DPAP2#6 I= 1.6 A; switch: A
    rotative encoder RE Encoder Cable: S/ CC133 VME Card   #Gr A / Ch 1
Detector Xtal-translation detF stepper motor STM Motor Cable: C/0138 VPAP RPAP8#1 Current Nor/Boost
    Limit   Limit Cable: C/0140 #M6/0 DPAP2#6 I= 1.6 A; switch: A
    Rotative encoder RE Encoder Cable: S/ CC133 VME Card   #Gr A / Ch 1
Photo of cryostream system
 
14 - Sample cryo cooling system  
OXFORD Cryo cooler X-spy OXFORD Controler RS232 Cable Serial line : C/0087 RS232 (Temp, status Â…) to VME IBAM    
        IBAM VME card #ch: I6      
LN2Level X-spy LN2 Level gauge AIN Cable LN2 level : S/0088 STB50 & ICV150 VME card #ch:8    
SuperDryerOut1   Super Dryer Controler DIN ICV196 VME card & STB116 Relay IN Chanel ICV196 Super drier 1: 42    
        Cable Super dryer 1 : S/0089      
SuperDryerOut2   Super Dryer Controler DIN ICV196 VME card & STB116 Relay IN Chanel ICV196 Super drier 2: 43    
        Cable Super dryer 2: S/0089      
NozleDryerFlow   Dry Air low meter switch DIN ICV196 VME card & STB116 Relay IN Chanel ICV196 Flow meter: 41    
        Cable Flow meter: S/0090      

Last Updated on 29/03/01

 

By Sylvain Labouré

European Synchrotron Radiation Facility