Determining if there is a problem with the Mar CCD temperature sensor
last modified
18-10-2010 14:37
mar detector with cover off, showing location of temp sensorEMBL/ESRF staff ONLY
- Shut down MAR software
- Dry off any condensation on the cooling inlet or outlet mesh tubes
- Turn off Mar controller
- Turn off main power (upper right hand side of mar rack
- Disconnect the main power cable to the back of the mar
- Remove cover from the back of the mar
- Now look to make sure the temp sensor is not touching anything
- Reconnect the power cable
- Turn on main power to mar rack
- turn on power to mar controller
- Are the fans working?
- Go through the shut down sequence again
- Replace cover
- Re-plug everything
- Start up controller
- Turn on main rack switch
- Turn on mar software