DPAP 2 setting 1
When you want to
configure a DPAP module for use with a stepper motor, you must :
- know the phase
current
- know the positive
direction
Current settings
:
Current
Table for DPAP2
Current
Table for DPAP1
You can adjust
the phase current by turning the 3 rotary switches:
Normal : normal
phase current
Boost : phase current
during the boost cycle
Off : holding current
(phase current at rest)
Switch setting
:
If you have the
wrong polarity for the limit switches (both limits led are on), you must
change the two jumpers
W2 and W3.
W2 :
Polarity change for the CW limit switch
W3 :
Polarity change for the CCW limit switch
Motor direction :
When you move the motor in the positive direction, you must hit the CW hard
limit.
If not, you have to change the jumper W15.
If the positive direction, according to the limit, is not correct, you must
change
the wiring in the motor connector itself.
Continuity Wiring Control :
If the led C, on the front panel, is on, change the jumper W4.