Insertion Devices
We have 3 insertion devices on ID19:
- 2 undulators: U32 and U44
- 1 wiggler
How to use our undulators (U32 and U44)
1) open gaps of both undulators: ~295.99 mm
- in Insertion Device Control application (on small monitor near the printer) fill corresponding "set" columns and then press "enter"
2) set diamond window for undulators
(ring of 5 mm diameter mounted on the first attenuator axis: max. beam size = 25 mm)
- in SPEC application NATT:
attshow
(semi graphic showing all absorber axes and their state)
- 104 (set fourth absorber/window of the first axis into
the beam, which corresponds to the small diamond for undulators)
3) close wiggler gap to ~ 50 mm (always in ID Control application)
4) choose desired photon energy E and send monochromator to it
- for a "convenient" energy consult the tables with energies of various harmonics as a function of the gap of two undulators, note which undulator at which gap you desired- in SPEC application VMONO: moveE E
5) check monochromatic beam with- fluorescent screen in blue/white box: in SPEC application HMONO: mv scr 0
- Laue camera: need to be put manually in the position in the experimental hutch where beam is expected (for example on hdiffr)
- open primary slits horizontal gap pshg more than 5 mm up to you can see on fluorescent screen or Laue camera the round borders of the diamond window
- in SPEC application SSLITS: mv pshg 7 or in the graphical primary slits application set pshg to 7 and enter
- if it is not the case, don’t close the gap of any undulator and consult a specialist (local contact)
- vacuum values should not exceed 10-8 mbar, you find them in graphical general beam line application called Vacuum
- temperatures should not exceed 50°C, they are in graphical application called Signal Monitoring
-
in the case of worrying values:
- a) try to use the smaller
beam possible (close primary slits)
b) change from U44 to U32
c) close Front-End and consult a specialist (as local contact)
9) close ONE undulator gap to desired value, chosen in item 4 (always in ID control application)
10) perform energy scan (be sure that monitor is operational)
-
in SPEC application VMONO
(example if E=10.5 keV desired):
nrj_scan 10 11 10 1
-
nrj_scan minE maxE_of_scan_interval number_of_steps counting_time_per_point
- in SPEC application VMONO: moveE maximum_E
- in SPEC application HMONO: mv scr -90
13) you are ready to work !